Stapling and Energy Devices for Endoscopic Surgery

Chapter 11 Stapling and Energy Devices for Endoscopic Surgery



Endoscopic procedures frequently involve tissue manipulation and dissection so that the target tissue can be exposed. Hemorrhage is not uncommon in such maneuvers. Effectively dealing with intraoperative hemorrhage is a prerequisite to successful completion of the endoscopic procedure. Special techniques and equipment exist that facilitate hemostasis in a limited space, such as that encountered in minimally invasive surgery. If hemorrhage is not controlled, conversion to an open procedure may be required. The following sections will explore the theory, application, and limitations of endomechanical and energy-based devices when used in minimally invasive surgery. In so doing, I seek to help the endoscopic surgeon make the best possible choices in utilizing these technologies.


Aside from ligatures and pretied suture loops, surgical stapling devices are the fundamental means of arresting hemorrhage during endoscopy. These vary in complexity from single-fire clip appliers to multifire endostaplers. Disposable loading units (DLUs) of several widths and staple leg lengths are available to accommodate variable tissue thickness and trocar sizes. Endoscopic gastrointestinal anastomosis (Endo GIA) devices include an integrated knife blade, which allows division between triple-staggered rows of staple lines during the firing cycle.


The past 10 years have seen significant advances in surgical devices geared toward operative dissection and hemostasis. Advanced electrosurgical generators, ultrasonic scalpels, and surgical lasers are examples of such. Although they exert their effects via different physical principles, they are best viewed simply as different means of delivering energy to tissue. Tissue effects are ultimately mediated by thermal changes, which are predictable despite the operative method (Figure 11-1). No one particular system is inherently superior to the other. Each has its advantages and disadvantages for a given procedure. Unfortunately, the choice of an energy modality is frequently more a function of surgeon familiarity and training rather than an informed decision.




Endosurgical Clips and Staplers


Endosurgical clip appliers were developed in 1990 and were used extensively in humans to perform laparoscopic cholecystectomy, one of the first major laparoscopic treatments in people. Surgical clip appliers have a place in every endoscopist’s armamentarium. A single clip application may negate the need to convert to an open procedure. Modern clip appliers can be precocked before placement, which allows the clip to be advanced proximal or distal along a tubular structure before closure. The distal tips of the clip are closed first, which prevents tissue from being squeezed laterally beyond the clip margins before closure. Some form of clip counter is visible on the instrument and ensures sufficient clip numbers to complete the procedure (Figure 11-2).



Failures attributable to clip application are usually a function of two factors: (1) inclusion of too much tissue within the clip and (2) clip slippage. The first error can be prevented by selection of the appropriate clip configuration and size. C-shaped clips that capture the tubular structure by closing at the ends first are generally most effective. Clip size is a function of shaft diameter, and 5- and 10-mm versions are available. Clip slippage can be minimized by application without undue tension on the structure to be clipped. Tissue diameter changes with the release of tension and can cause a clip to migrate away from the original application site.


Multifire endostapler cartridges (DLUs) come in a variety of lengths and configurations (Figure 11-3). The staples themselves are composed of titanium and are formed into a B shape as they are applied against the cartridge anvil during the firing cycle. The B shape allows microvascular perfusion to the staple line, which prevents necrosis that could lead to delayed hemorrhage or leakage. Available options include an integrated cutting blade as well as a roticulating hinge. Roticulation is defined as the ability to simultaneously rotate and articulate and requires the appropriate endostapler as well as stapler cartridge. Note that the use of a roticulating as opposed to a simple straight cartridge adds approximately 1.5 cm of length to the DLU, which can be an important consideration when working in a tight space, such as the chest.



Numerous staple leg and cartridge lengths allow application of hemostatic and pneumostatic staple lines across parenchymal, bronchial, and vascular pedicles of varying thicknesses. In thick parenchymal tissue applications, such as the liver, the staple lines may be placed in tiers to effectively seal the cut edges in a stepwise fashion.


It is important to be attentive to staple closure widths when stapling devices are used. If the tissue to be stapled is either thinner or thicker than the closed width of the staple, then additional measures need be taken to prevent hemorrhage or leakage. These measures include suture over-sew and/or application of ligating loops, clips, and energy-based technologies.



Electrosurgical Generators


The first electrosurgical generator was developed by Harvey Cushing and William Bovie in the 1920s. Early electrosurgical generators that followed from their original “Bovie” were deficient primarily in two areas: patient safety and consistent tissue effect. Advances in computer and electronics technologies in the past 20 years have not only addressed these issues but also established electrosurgery as the preferred standard in many operating theaters.



Theory


Electricity is defined as the flow of electrons from a source (+ electrode) to another electrode (− electrode) or ground. The flow of electricity along a circuit is best described by



image



where V = voltage; I = current; and R = resistance to flow


This can be conceptually illustrated by water flowing through a garden hose. For water to flow through the hose, there must be driving pressure to overcome the resistance to flow imposed by the hose diameter. Voltage is that driving force, and resistance relates to the hose diameter. Current represents the total amount of water per unit time, which is dependent on the driving pressure (voltage) and the resistance to flow due to the hose diameter. When applied to a biological system, resistance is referred to as impedance and is a function of blood supply and tissue composition. Furthermore, as electricity is applied to tissue, its impedance is continually changing as it undergoes desiccation.


Household electricity is supplied as 110-volt, alternating current at 60 Hz. That is, the current is supplied as a sine wave, which changes polarity from positive to negative at 60 cycles per second. The neuromuscular system becomes refractory to electrical stimulation beyond a frequency of 100,000 Hz. For this reason, modern electrosurgical generators utilize frequencies in the range of 350,000 to 500,000 Hz, which falls within the medium radiofrequency electromagnetic spectrum (Figure 11-4). Some units may go as high as 3 to 4 megahertz (MHz) and deserve special mention (Figure 11-5). These units are commonly referred to as radiosurgery units, but this terminology is not only confusing but incorrect. By medical convention, the term radiosurgery applies to the use of a targeted beam of ionizing radiation used primarily for cancer ablation. The correct terminology for high-frequency electrosurgical application as referenced here is radio wave radiosurgery (RWRS). This type of system will be discussed separately in a subsequent section.




When direct current is allowed to flow through a wire connecting the positive and negative poles of an electrical circuit, the wire itself becomes hot. When this heated wire loop is brought into contact with tissue, the heat of the wire exerts a tissue effect proportional to its magnitude. This is referred to as electrocautery. This is in contrast with the system that is utilized in modern electrosurgical generators, whereby alternating current from an active electrode is allowed to pass through the patient’s body to a passive or dispersive electrode, and then back to the generator box. Although seemingly a minor point, it is important to realize this distinction, as the literature is fraught with references to electrocautery when electrosurgery as applied by an electrosurgical unit (ESU) is what is actually meant.


Power is defined by the following relationship:



image



where P = power; V = voltage; and I = current


In the context of an ESU, the selected power setting is a function of voltage and current. Voltage is a primary determinant of tissue effect and is very much a function of the waveform delivered by the generator.



Waveforms


Electrosurgical waveforms consist of essentially three types: cutting, coagulating, and blended (Figure 11-6). The cutting and blended waveforms are continuous and use less peak voltage than a coagulating waveform at the same power setting. Higher peak voltages result in greater lateral thermal damage to the target tissue. This is clinically apparent when tissue is incised with three different waveforms applied at the same power setting (Figure 11-7). For this reason, it is important for the surgeon to have a fundamental understanding of the physics of electrosurgery rather than relying exclusively on the selected waveform to exert its designated effect. Another clinically relevant example lies in the application of energy to a hemostat to achieve coagulation of a vessel. Intuitively, one would expect the operator to apply coagulating energy to the hemostat, when in fact this is not recommended practice. Cutting energy should be applied to the hemostat, as lesser voltage is utilized with less thermal spread. In fact, cutting energy can be used to effect hemostasis and is often underutilized in this capacity. Cutting energy is capable of producing deeper hemostasis than that achieved with coagulating energy. The reason for this lies in the fact that coagulating energy produces greater thermal damage, resulting in rapid buildup of tissue resistance from char accumulation at the tissue–electrode interface. As tissue resistance increases, greater power is needed to penetrate deeper tissue. With cutting energy, tissue heats up more quickly, steam forms as cells explode, and there is less char accumulation. Energy is able to penetrate deeper into the target tissue without stalling at the point of tissue damage.




Many surgeons prefer to use blended energy for nearly all tissue applications, considering it to be a tradeoff between tissue damage and inadequate hemostasis. As noted in Figure 11-6, blended energy is delivered as an interrupted waveform, with varying degrees of on and off time. As the percent of on time is increased, the more rapidly tissue is desiccated and hemostasis is achieved.



Monopolar Systems


Monopolar electrosurgical systems consist of a generator box, an electrosurgical pencil, and a return electrode. The generator box allows operator selection of output waveform as well as power. Electrical energy must flow from the grounded generator box to the electrosurgical pencil tip, through the patient to the attached dispersive electrode, and back to the generator. The electrosurgical pencil is referred to as the active electrode and the patient return pad as the passive (or dispersive) electrode. It is important that concentrated energy at the active electrode be returned to the patient over a wide contact area so as to reduce the chance of current concentration and patient burning. It is in this area that significant advances have been made to ensure patient safety.


Early monopolar systems utilized metal grounding plates to collect energy passing through the patient and return it to the generator. The weakness with these systems lay in the inconsistent contact interface between the patient and the rigid plate, occasionally resulting in unanticipated patient burns. The advent of malleable grounding pads alleviated much of this problem, but even these could lose adhesion during a procedure, which resulted in areas of increased current density and thermal damage. This problem was remedied when contact quality monitoring (CQM) circuitry was built into the generator. With this system, the generator was programmed to shut off if patient pad impedance fell outside of prescribed limits. A further improvement to this system came in the form of return electrode monitoring (REM), wherein an interrogation signal is sent from the generator to a split patient return pad. Again, the generator is programmed to cease operation if the return signal is outside programmed limits.


The generator box (ESU) converts standard 60-Hz, 110-volt current into an operator-selected waveform and power delivered at 300 to 500 KHz. The first monopolar generators were totally dependent on the operator with respect to ultimate tissue effect. Variations in time of application as well as inconsistent power delivery made for inconsistent results. With the advent of computer-modulated circuitry, however, huge advances were made to allow nearly real-time monitoring of tissue impedance. Modern generators sense tissue impedance at 200 times per second and alter voltage and current to deliver consistent power over a wide range of tissue impedances. The end result is much more consistent tissue effects (Figure 11-8).


Stay updated, free articles. Join our Telegram channel

Sep 10, 2016 | Posted by in SMALL ANIMAL | Comments Off on Stapling and Energy Devices for Endoscopic Surgery

Full access? Get Clinical Tree

Get Clinical Tree app for offline access