Chapter 66 The authors believe the quality-control program can be somewhat simplified for the veterinary practice facility. The most practical way to integrate quality control is to include it as part of daily start-up procedures, typically in the morning when the workday is begun. Most systems are in a stand-by mode when personnel arrive in the morning. Most benchtop units have a wake-up or start-up procedure. Analysis of a quality-control material at the conclusion of the start-up procedure verifies that the system is working properly and is ready to analyze patient samples. The procedure involves handling the quality-control material according to manufacturer directions. The material is accompanied by target values determined by repetitive reference procedures. These are expressed as a range for each measurement. Results from the analyzer are expected to fall within this range on a daily basis when all elements are functioning properly. The software of most benchtop instrument systems can store control range values and present control analysis results for inspection. The system also should be able to store control data in a way that can be retrieved for trend analysis. Additional background on quality-control procedures is available (Weiser and Thrall, 2007). More advanced treatment of quality-control data is discussed later in the chapter. • Sample evaluation. Occasionally, results may be obtained that do not fit the clinical picture or preconceived expectations. When this happens, the sample and sample handling should be evaluated. In blood chemistry analyzers, the presence of interfering substances such as those associated with lipemia or hemolysis, the presence of fibrin clots interfering with sample loading, and improper anticoagulant use are considerations for evaluation. In hematologic testing, the adequacy of sample mixing and the presence of microclots in the sample are considerations, as are hemolysis or lipemia. Instrumentation suppliers should provide guidelines for this evaluation, and their technical support personnel may provide additional help in the evaluation when needed. One should keep in mind that it may be appropriate to repeat the analysis when an unexplained aberrant value occurs. • Use of quality-control material outside the routine cycle. In addition to performing regular analysis of quality-control material at the beginning of each day, occasionally it is appropriate to check system function with quality-control material whenever it is suspected that analytical failure may be occurring. • Blood film examination. It is important to note that the capability of hematology analyzers to produce a differential distribution is not intended to replace blood film review of leukocytes. At best, it is intended to detect a reasonably normal hemogram. Furthermore, there is no good quality-control material for the differential analysis. Hematology analyzers cannot detect abnormalities such as left-shifted cells, leukemic cells, nucleated erythrocytes, and other abnormal cell types. When there is leukopenia, leukocytosis, an abnormality in the differential distribution, or any abnormality in the instrument’s histogram or cytogram display, a blood film scan should be performed to corroborate the generated differential data as well as to evaluate erythrocyte morphology and screen for hemoparasites. A differential count should be performed by microscopy whenever a discrepancy or abnormal cells are detected by the scan. • Evaluation of mean corpuscular hemoglobin concentration (MCHC). The relationship between hemoglobin concentration (Hb) and hematocrit (Hct) is a physiologic constant. These two values are used to calculate the MCHC: MCHC = (Hb × 100)/Hct. The MCHC is typically 32 to 36 g/dl, with minor variation among instrument systems. Because the two primary measurements are performed independently by taking separate measurements on two separate dilutions, they can be used to corroborate each other. Certain pathologic conditions, including iron deficiency and marked regeneration, may cause a minimal decrease in MCHC (e.g., to 30 g/dl). Certain sample abnormalities may variably increase MCHC by artifact, including marked lipemia, sample hemolysis, pathologic red blood cell (RBC) agglutination, and marked numbers of Heinz bodies. Occasionally an Hct value does not meet clinical expectations. There are two readily available means to evaluate this. One is to corroborate the Hct value with the Hb concentration. This is most easily done by examination of the MCHC value; if it is in the physiologic reference interval, the Hct and Hb are confirming each other. A nonsensically low MCHC value indicates an analytical malfunction. A very high MCHC value may indicate the same after sample abnormalities such as marked lipemia, marked sample hemolysis, prominent RBC agglutination, and marked numbers of Heinz bodes are ruled out. The second method is to compare the instrument-derived Hct with a manually obtained packed cell volume. Results for the two methods should match within a reasonable limit, provided the collection tube is filled properly. This limit may vary slightly with the instrument and the patient’s disorder. Failure of the results to match can localize the source of the nonsense MCHC values to the RBC counting and sizing. In addition, the centrifuged microhematocrit tube provides a simple means of examining the supernatant for hemolysis and lipemia. If nonsense values are obtained repeatedly, analyzing the quality-control material or contacting technical support is warranted.
Quality Control for the In-Clinic Laboratory
Background
Implementations for Chemistry and Hematology Analyzers
Adjunct Procedures That Supplement Routine Use of Quality-Control Material
Stay updated, free articles. Join our Telegram channel
Full access? Get Clinical Tree
Quality Control for the In-Clinic Laboratory
Only gold members can continue reading. Log In or Register a > to continue