Electrophysical Agents in Physiotherapy

Chapter 94Electrophysical Agents in Physiotherapy



Electrotherapy has been a component of physiotherapy practice since the early days, but its delivery has changed remarkably and continues to do so. All electrotherapy modalities involve the introduction of some physical energy, which brings about one or more physiological changes that are used for therapeutic benefit. To appropriately select the most suitable modality, it is necessary to:









Therapeutic Ultrasound


Ultrasound (US) is a form of mechanical energy, not electrical energy, and therefore, strictly speaking, is not electrotherapy but falls into the group of electrophysical agents.1 Mechanical vibration at increasing frequencies is known as sound energy. In children and young adults, the normal audible sound range is 16 Hz to 15,000 to 20,000 Hz. Higher frequencies are known as US. The frequencies used in therapy are typically between 1.0 and 3.0 MHz (1 MHz = 1 million cycles/sec).


Sound waves are longitudinal waves consisting of areas of compression and rarefaction. When exposed to a sound wave, particles of a material oscillate about a fixed point rather than move with the wave itself. Any increase in the molecular vibration in a tissue can result in heat generation; thus US can be used to produce thermal changes in the tissues, although current therapeutic usage does not focus on this phenomenon.1,2 The vibration of the tissues may also have nonthermal effects. As the US wave passes through the tissue, the energy levels within the wave decrease as energy is transferred to the tissues.1





Ultrasound Transmission through the Tissues


Tissues present impedance to the passage of sound waves. The specific impedance of a tissue is determined by its density and elasticity. For the maximal transmission of energy from one tissue to another, the impedance of the two tissues needs to be as similar as possible. An air gap between the generator and the skin will result in the majority of the US energy being reflected rather than transmitted to the underlying tissues.


Coupling media, water, various oils, creams, and gels, are used to bridge the air gap. Ideally, the coupling medium should be fluid so as to fill all available spaces, be relatively viscous so that it stays in place, have an impedance appropriate to the media it connects, and allow transmission of US with minimal absorption, attenuation, or disturbance. At present, the gel-based media appear to be preferable to the oils and creams. Water is a good medium and can be used as an alternative. The addition of active agents (e.g., antiinflammatory drugs) to the gel is widely practiced but remains incompletely researched.


Studies3,4 have considered the effect of animal hair on the transmission of US to the underlying tissue and report that best penetration is achieved by clipping the hair. In addition to the reflection that occurs at a boundary because of differences in impedance, there will also be some refraction if the wave does not strike the boundary surface at 90 degrees. Essentially, the direction of the US beam through the second medium will not be the same as its path through the original medium; its pathway is angled. The treatment head is ideally placed perpendicular to the skin surface (i.e., at 90 degrees). If the treatment head is at an angle of 15 degrees or more from the perpendicular to the plane of the skin surface, the majority of the US beam will travel through the dermal tissues (i.e., parallel to the plane of the skin surface) rather than penetrate the tissues as would be expected.







Therapeutic Effects of Ultrasound


The therapeutic effects of US are influenced by the treatment parameters chosen and are commonly divided into thermal and nonthermal.




Nonthermal


The nonthermal effects of US are from cavitation and acoustic streaming.7,12 Cavitation, the formation of gas-filled voids within tissues and body fluids, occurs in two types—stable and unstable—which have different effects. Stable cavitation occurs at therapeutic doses of US and is the formation and growth of dissolved gas bubble accumulation. The “cavity” acts to enhance the acoustic streaming phenomena. Unstable (transient) cavitation is the formation of bubbles at the low pressure part of the US cycle. These bubbles then collapse very quickly, releasing a large amount of energy that is detrimental to tissue viability. However, this does not occur at therapeutic levels if good technique is used.


Acoustic streaming is a small-scale eddying of fluids near a vibrating structure such as cell membranes and the surface of a stable cavitation gas bubble and affects diffusion rates and membrane permeability.11 Sodium ion permeability is altered, resulting in changes in the cell membrane potential. Calcium ion transport is modified, which in turn leads to an alteration in the enzyme and cellular secretions.


The result of the combined effects of stable cavitation and acoustic streaming is that the cell membrane becomes “excited” (up-regulates), thus increasing the activity levels of the whole cell. The US energy acts as a trigger for this process, but it is the increased cellular activity that is in effect responsible for the therapeutic benefits of the modality.1,2,13


Some US machines offer variable time: typical pulse ratios are 1 : 1 and 1 : 4. In 1 : 1 mode, the machine offers an output for 2 ms followed by 2 ms of rest. In 1 : 4 mode, the 2-ms output is followed by 8-ms rest period.



Ultrasound Application in Relation to Soft Tissue Repair


The process of tissue repair is a complex series of cascaded, chemically mediated events that lead to the production of scar tissue.



Stay updated, free articles. Join our Telegram channel

Jun 4, 2016 | Posted by in EQUINE MEDICINE | Comments Off on Electrophysical Agents in Physiotherapy

Full access? Get Clinical Tree

Get Clinical Tree app for offline access